Sens Actuators A Phys
December 2013
We present a passive, miniature check valve which can be manufactured using standard techniques ideal for low-cost, disposable systems used in medical devices and other applications. The body of the valve consists of a hollow cylindrical core, closed at one end, with a side port and a cylindrical elastomeric sleeve placed over the core body, covering the side port. The pressure required for initial opening of the valve, referred to as cracking pressure, can be adjusted, and depends predominantly on the valve core outer diameter, the sleeve inner diameter, the sleeve wall thickness, and the sleeve material's modulus of elasticity.
View Article and Find Full Text PDFInfectious disease diagnosis in point-of-care settings can be greatly improved through integrated, automated nucleic acid testing devices. We have developed an early prototype for a low-cost system which executes isothermal DNA amplification coupled to nucleic acid lateral flow (NALF) detection in a mesofluidic cartridge attached to a portable instrument. Fluid handling inside the cartridge is facilitated through one-way passive valves, flexible pouches, and electrolysis-driven pumps, which promotes a compact and inexpensive instrument design.
View Article and Find Full Text PDF