An on-chip type cation-exchange chromatography system with electrochemical detection of HbA(1c), which is one of the most important diabetes marker protein, was developed using ferrocene-conjugated anti-human hemoglobin (Hb) monoclonal antibody (FcAb). The FcAb was used as an electrochemical probe for the detection of each Hb. The system contains syringe pump, on-chip type cation-exchange column consisted of PDMS and cation-exchange resin beads, and a three-electrode flow-cell system.

View Article and Find Full Text PDF

An on-chip electrochemical flow immunoassay system for the detection of hemoglobin A1c (HbA1c) was developed using anti-human hemoglobin (Hb) IgG labeled with ferrocene monocarboxylic acid (Fc-COOH) and boronate-affinity chromatography. An on-chip column packed with boronate-activated agarose beads was used for the separation of HbA1c from both non-glycated Hb and free antibody. Anti-human Hb IgG conjugated to Fc-COOH (Fc-IgG) was used for the electrochemical detection of HbA1c.