Objective: Standards for myocardial single-photon emission computed tomography (SPECT) adapted for a Japanese population were not available. The purpose of this study was to create standard files approved by the Japanese Society of Nuclear Medicine and to make known the characteristics of the myocardial perfusion pattern of this population.
Methods: With the collaboration of nine hospitals, a total of 326 sets of exercise-rest myocardial perfusion images were accumulated from subjects with a low likelihood of cardiac diseases.
Improvements in image quality and quantitation measurement, and the addition of detailed anatomical structures are important topics for single-photon emission tomography (SPECT). The goal of this study was to develop a practical system enabling both nonuniform attenuation correction and image fusion of SPECT images by means of high-performance X-ray computed tomography (CT). A SPECT system and a helical X-ray CT system were placed next to each other and linked with Ethernet.