We compared plastic (polycarbonate) and high-quality glass support materials for gold-coated slides, when performing a model immunoassay against rabbit IgG using fluorescently labeled (AlexaFluor-647) anti-rabbit IgG, and detecting surface plasmon-coupled emission (SPCE) signals. Both, glass and plastic slides were simultaneously coated with a 48-nm layer of gold and protected with a 10-nm layer of silica. The maximum SPCE signal of AlexaFluor-647 was only two- to three-fold smaller on plastic slides than on glass slides.

View Article and Find Full Text PDF

We present a new approach for performing fluorescence immunoassay in whole blood using fluorescently labeled anti-rabbit immunoglobulin G (IgG) on a silver surface. This approach, which is based on surface plasmon-coupled emission (SPCE), provides increased sensitivity and substantial background reduction due to exclusive selection of the signal from the fluorophores located near a bioaffinity surface. This article describes the effect of an optically dense sample matrix, namely human whole blood and serum, on the intensity of the SPCE.

View Article and Find Full Text PDF

We used the Hamamatsu model PLP-01 picosecond light pulser as a 413-nm excitation light source for frequency-domain fluorescence measurements. In comparison with sync-pumped/cavity dumped/frequency-doubled dye lasers, the 413-nm PLP-01 shows a longer FWHM (40 ps), a similar pulse repetition rate (up to 10 MHz), much less output power at a fixed wavelength (0.44 mW peak, 220 nW maximum average power), but is less expensive, small-sized, and easy to handle.

View Article and Find Full Text PDF