Background: As several new tibial osteotomy plates recently appeared on the market, the aim of the present study was to compare mechanical static and fatigue strength of three newly designed plates with gold standard plates for the treatment of medial knee joint osteoarthritis.
Methods: Sixteen fourth-generation tibial bone composites underwent a medial open-wedge high tibial osteotomy (HTO) according to standard techniques, using five TomoFix standard plates, five PEEKPower plates and six iBalance implants. Static compression load to failure and load-controlled cyclic fatigue failure tests were performed.
Background. The objective of the present study was to compare mechanical strength and stability of the newly designed spacer plate with the gold standard plate for the treatment of medial knee joint osteoarthritis. Materials and Methods.
View Article and Find Full Text PDFKnee Surg Sports Traumatol Arthrosc
January 2013
Purpose: Biplanar open-wedge high tibial osteotomy (HTO) is thought to promote rapid bone healing due to the increased cancellous bone surface compared to other HTO techniques. However, precise data on the bone surface area and wedge volume resulting from both open- and closed-wedge HTO techniques remain unknown. We hypothesized that biplanar rather than uniplanar HTO better reflects the ideal geometrical requirements for bone healing, representing a large cancellous bone surface combined with a small wedge volume.
View Article and Find Full Text PDF