Among factors affecting the reliability of heart valve prostheses (HVP) the following ones are of a great importance: the temperature and the duration of soldering a big clamp of the HVP, the occurrence of nonmetal inclusions in the original material (the alloy 45KXBH), etc. Analysis is given to the factors causing the destruction of the HVP during its operation by contact effects, and the possibility of such failures is eliminated. The technique to test the soldering nondestructively is discussed.