The QuEChERS (quick, easy, cheap, effective, rugged, and safe) sample preparation method was modified to accommodate various cereal grain matrices (corn, oat, rice, and wheat) and provide good analytical results (recoveries in the range of 70-120% and RSDs <20%) for the majority of the target pesticides (about 180 analytes). The method consists of a 1 h shaking of a milled sample (2.5-5 g) in 20 mL of 1:1 (v/v) water/acetonitrile (or 25 mL of 1.