Purpose: To perform the final quality assurance of our fluoroscopic-based markerless tumor tracking for gated carbon-ion pencil beam scanning (C-PBS) radiotherapy using a rotating gantry system, we evaluated the geometrical accuracy and tumor tracking accuracy using a moving chest phantom with simulated respiration.
Methods: The positions of the dynamic flat panel detector (DFPD) and x-ray tube are subject to changes due to gantry sag. To compensate for this, we generated a geometrical calibration table (gantry flex map) in 15° gantry angle steps by the bundle adjustment method.