To develop and benchmark a novel 3D dose verification technique consisting of polymer gel dosimetry (PGD) with cone-beam-CT (CBCT) readout through a two-institution study. The technique has potential for wide and robust applicability through reliance on CBCT readout..
View Article and Find Full Text PDFBackground: Sophisticated modern radiation therapy treatments require comprehensive validation in 3D.
Purpose: Investigation and characterization of a novel 3D dosimetry system consisting of ClearView radiochromic gel dosimeters (commercially available from Modus Inc) and an in-house telecentric optical CT scanner DLOS (the Duke Large Field of View Optical-CT Scanner).
Methods: Spectrophotometry measurements were made on small volumes of ClearView gel irradiated with 6X photon doses up to 40 Gy to determine linearity and temporal stability of dose response.