We are presenting a nightmare case of successful retrieval of the broken distal tip of a multipurpose catheter which embolised into the left pulmonary artery (LPA) during atrial septal defect device closure. During the course of snaring it initially slipped once in the LPA and subsequently while snaring it into the delivery sheath it slipped again in the inferior vena cava and yet again embolised into the LPA!! Subsequent attempt was successful but the distal fragment itself broke into two parts while being pulled inside the delivery sheath. Luckily the fragments did not embolise again as they were stuck at the tip of the sheath.
View Article and Find Full Text PDFSeven-year-old boy presented with pain in right lower limb and abdomen after a fall from a bullock cart. He was referred for management of pseudoaneurysm of the right common iliac artery. After CT angiogram, the vascular surgeon opined that child was not suitable for surgical patch or graft or endovascular stenting as there was no landing zone for the stent.