The data presented here was produced as part of an evaluation of the performance of the CoaguChek XS point-of-care coagulation analyzer, which is discussed in the research article "POCT PT INR - Is it adequate for Patient Care? A Comparison of the Roche Coaguchek XS vs. Stago Star vs. Siemens BCS in Patients Routinely Seen in an Anticoagulation Clinic" (Baker et al.

View Article and Find Full Text PDF

Background: In this study we examined the difference in patient INR values as measured by the POCT CoaguChek XS device and central laboratory Stago Evolution and Siemens BCS XP analyzers.

Methods: This study composed of 100 warfarin therapy patients and 20 coagulation normal subjects, showed that the difference between the POCT and clinical laboratory values increased with increasing INR and was exacerbated by the use of different thromboplastin reagents by the POCT and central lab.

Results: The CoaguChek XS and on-site Stago analyzers which used human recombinant (ISI=1.

View Article and Find Full Text PDF

To assess the impact of preanalytical variables of time and temperature on prothrombin time (PT), activated partial thromboplastin time (aPTT), dilute Russell viper venom time (DRVVT), activated protein C resistance (APCR), and d-dimer, samples from 23 healthy individuals and 18 patients having coagulopathy with known abnormal PT and aPTT were collected. Plasma from each individual was separately pooled and aliquoted; the first 2 aliquots were stored at room temperature then analyzed at 2 hours (baseline) and 4 hours postcollection. The remaining aliquots were stored at -20°C and thawed for analysis at 48 hours, 1, and 2 weeks.