The information obtainable from a radiograph depends on the viewing conditions under which the radiograph is viewed. All the viewing conditions can be defined in terms of scattered light. The amount of information taken from a radiograph with S = 1000 (S = speed class) and perfect conditions is the same as that from a radiography with S = 100 viewed under poor conditions.

View Article and Find Full Text PDF

It is shown that the number of contrasts which can be differentiated on a radiograph depends on the detail size, the noise, and the viewing conditions. The minimum signal-to-noise ratio (SNR) for a perception approach is equivalently described by the SGA value. This testing method was first developed by De Belder and Bollen to examine the visibility of low contrast details.

View Article and Find Full Text PDF

The present article describes the circumstances concerning the use of testing aids such as sensitometers with one-sided exposure. It is shown which phenomena must be considered if radiographic films coated on both sides are exposed with a) standard pocket sensitometers (one-sided exposure), b) lab sensitometers (double-sided exposure to ANSI Ph 2.9 [1964]), c) x-radiation in the cassette, intensifying screen and film system (to DIN 6867 T 1).