Background: Dose area product in water (DAP) in small fields relies on the use of detectors with a sensitive area larger than the irradiation field. This quantity has recently been used to establish primary standards down to 5 mm field size, with an uncertainty smaller than 0.7%.
View Article and Find Full Text PDFPurpose: To present primary standards establishment in terms of Dose Area Product (DAP) for small field sizes.
Methods: A large section graphite calorimeter and two plane-parallel ionization chambers were designed and built in-house. These chambers were calibrated in a 6MV FFF beam at the maximum dose rate of 1400 UM/min for fields defined by specifically designed circular collimators of 5, 7.