Introduction: Falsely elevated synovial white blood cell (WBC) counts using automated hematology analyzers have been reported particularly in the setting of joint arthroplasty. We evaluated the implementation of a laboratory workflow based on Sysmex XN-1000-automated cell counting and scattergram interpretation.
Methods: WBC and differential were measured for 76 synovial fluid samples (29 native joints and 47 with joint arthroplasties) with Sysmex XN-1000 and manual methods.