Publications by authors named "Juha-Pekka Nuutine"

Objective: To compare the biocompatibility properties of a new braided biodegradable self-reinforced poly-L-lactic acid (SR-PLLA) urethral stent to the former spiral biodegradable SR-PLLA stent and the stainless steel stent in a rabbit model.

Materials And Methods: In all, 54 male New Zealand White rabbits were anaesthetized and stents inserted into the prostatic urethra, three of each kind for each sample time. The rabbits were killed after 1, 3, 6, 9, 12 or 15 months and light microscopy and scanning electron microscopy used to analyse the effects.

View Article and Find Full Text PDF