Publications by authors named "Juergen Dax"

So far, surface electromyography (sEMG) has been the method of choice to detect and evaluate muscle fatigue. However, recent advancements in non-cryogenic quantum sensors, such as optically pumped magnetometers (OPMs), enable interesting possibilities to flexibly record biomagnetic signals. Yet, a magnetomyographic investigation of muscular fatigue is still missing.

View Article and Find Full Text PDF

Objective: This proof-of-principle-study evaluated the extent to which spontaneous activity (SA) of the muscle can be detected via non-invasive magnetomyography (MMG) with optically pumped magnetometers (OPM).

Methods: Five patients, who together exhibited all forms of SA (fibrillations, positive sharp waves, fasciculations, myotonic discharges, complex-repetitive discharges) with conventional needle electromyography (EMG), were studied by OPM-MMG and simultaneous surface EMG (sEMG) while at rest, during light muscle activation, and when a muscle stretch reflex was elicited. Three healthy subjects were measured as controls.

View Article and Find Full Text PDF