To use the open-source Monte Carlo (MC) software calculations for TPS monitor unit verification of VMAT plans, delivered with the Varian TrueBeam linear accelerator, and compare the results with a commercial software product, following the guidelines set in AAPM Task Group 219. The TrueBeam is modeled in EGSnrc using the Varian-provided phase-space files. Thirteen VMAT TrueBeam treatment plans representing various anatomical regions were evaluated, comprising 37 treatment arcs.
View Article and Find Full Text PDFWe have developed a fast and accurate in-house Monte Carlo (MC) secondary monitor unit (MU) check method, based on the EGSnrc system, for independent verification of volumetric modulated arc therapy (VMAT) treatment planning system dose calculations, in accordance with TG-114 recommendations. For a VMAT treatment plan created for a Varian Trilogy linac, DICOM information was exported from Eclipse. An open-source platform was used to generate input files for dose calculations using the EGSnrc framework.