Background: Independent verification of the dose delivered by complex radiotherapy can be performed by electronic portal imaging device (EPID) dosimetry. This paper presents 5-yr EPID in vivo dosimetry (IVD) data obtained using the Dosimetry Check (DC) software on a large cohort including breast, lung, prostate, and head and neck (H&N) cancer patients.
Material And Methods: The difference between in vivo dose measurements obtained by DC and point doses calculated by the Eclipse treatment planning system was obtained on 3795 radiotherapy patients treated with volumetric modulated arc therapy (VMAT) (n = 842) and three-dimensional conformal radiotherapy (3DCRT) (n = 2953) at 6, 10, and 15 MV.
Purpose: Spatially fractionated radiotherapy is a strategy to overcome the main limitation of radiotherapy, i.e., the restrained normal tissue tolerances.
View Article and Find Full Text PDFPurpose: The primary aim of this study was to determine correction factors, kQclin,Qmsrfclin,fmsr for a PTW-31016 ionization chamber on field sizes from 0.5 cm × 0.5 cm to 2 cm × 2 cm for both flattened (FF) and flattened filter-free (FFF) beams produced in a TrueBeam clinical accelerator.