Purpose: The aim of this work is to investigate the effects of immersing EBT3 radiochromic film in water and to evaluate its contribution to the total uncertainty in dose determination.
Materials And Methods: We used 3 cm × 3 cm EBT3 radiochromic films irradiated in the range of 0-70 Gy to study the impact of water immersion on the change in net optical density. These films were placed in a water container for a period of 24 h.
J Appl Clin Med Phys
March 2018
The Canadian Organization of Medical Physicists (COMP), in close partnership with the Canadian Partnership for Quality Radiotherapy (CPQR) has developed a series of Technical Quality Control (TQC) guidelines for radiation treatment equipment. These guidelines outline the performance objectives that equipment should meet in order to ensure an acceptable level of radiation treatment quality. The TQC guidelines have been rigorously reviewed and field tested in a variety of Canadian radiation treatment facilities.
View Article and Find Full Text PDFPurpose: To investigate empirically the energy dependence of the detector response of two in vivo luminescence detectors, LiF:Mg,Cu,P (MCP-N) high-sensitivity TLDs and Al O :C OSLDs, in the 40-300-kVp energy range in the context of in vivo surface dose measurement. As these detectors become more prevalent in clinical and preclinical in vivo measurements, knowledge of the variation in the empirical dependence of the measured response of these detectors across a wide spectrum of beam qualities is important.
Method: We characterized a large range of beam qualities of three different kilovoltage x-ray units: an Xstrahl 300 Orthovoltage unit, a Precision x-Ray X-RAD 320ix biological irradiator, and a Varian On-Board Imaging x-ray unit.
Radiochromic film has become an important tool to verify dose distributions for intensity-modulated radiotherapy (IMRT) and quality assurance (QA) procedures. A new radiochromic film model, EBT3, has recently become available, whose composition and thickness of the sensitive layer are the same as those of previous EBT2 films. However, a matte polyester layer was added to EBT3 to prevent the formation of Newton's rings.
View Article and Find Full Text PDFThis is a proof of principle study on an algorithm for optimizing external beam radiotherapy in terms of both photon beamlet energy and fluence. This simultaneous beamlet energy and fluence optimization is denoted modulated photon radiotherapy (XMRT). XMRT is compared with single-energy intensity modulated radiotherapy (IMRT) for five clinically relevant test geometries to determine whether treating beamlet energy as a decision variable improves the dose distributions.
View Article and Find Full Text PDFPurpose: Total body irradiation (TBI) techniques aim to deliver a uniform radiation dose to a patient with an irregular body contour and a heterogeneous density distribution to within +/-10% of the prescribed dose. In the current article, the authors present a novel, aperture modulated, translating bed TBI (AMTBI) technique that produces a high degree of dose uniformity throughout the entire patient.
Methods: The radiation beam is dynamically shaped in two dimensions using a multileaf collimator (MLC).
During the recommissioning of a Philips RT-250 kilovoltage X-ray unit, unexpected output variations with tube head rotation (cross-plane) and tube head tilt (in-plane) were observed. The measured output showed an increase of up to 7.3% relative to the neutral position (0? in-plane and 0? cross-plane) over the possible range of angles of in-plane rotation for 75 kVp (half-value layer, HVL = 1.
View Article and Find Full Text PDF