We present the design and procedures for implementing a parallel optical coherence tomography (POCT) imaging system that can be adapted to an endoscopic format. The POCT system consists of a single mode fiber (SMF) array with multiple reduced diameter (15 microm) SMFs in the sample arm with 15 microm center spacing between fibers. The size of the array determines the size of the transverse imaging field.

View Article and Find Full Text PDF

The design and fabrication procedures for implementing a high-density (16-microm center spacing) single-mode fiber (SMF) array are described. The specific application for this array is a parallel optical coherence tomography system for endoscopic imaging. We obtained fiber elements by etching standard single-mode SMF-28 fibers to a diameter of 14-15 microm.