Purpose: To investigate the dependency of dose-volume histogram (DVH) behavior and precision on underlying discretization using shapes and dose distributions with known analytical DVHs for five commercial DVH calculators.
Methods: DVHs and summary metrics were extracted from all five systems using synthetic cone and cylinder objects for which the true volume and DVH curves were known. Trends in the curves and metrics were explored by varying the underlying voxelization of the CT image, structure set, and dose grid as well by varying the geometry of the structure and direction of a linear dose gradient.
Purpose: To assess the dosimetric accuracy and energy dependence of the new EBT-eXtended Dose (XD) Gafchromic film and to compare the lateral response artifact (LRA) between EBT-XD and EBT3 film.
Methods: EBT3 and EBT-XD calibration curves were created by exposing films to known doses from 0 to 3000 cGy using a 6 MV beam. To assess the accuracy and dynamic range of EBT-XD, a 60° enhanced dynamic wedge (EDW) was used to deliver a dose range of approximately 200-2900 cGy.