Purpose: To compare photon-counting CT (PCCT) and multi-detector CT (MDCT) for visualization of temporal bone anatomic structures.

Methods: Thirty-six exams of temporal bones without pathology were collected from consecutive patients on a MDCT, and another 35 exams on a PCCT scanner. Two radiologists independently scored visibility of 14 structures for the MDCT and PCCT dataset, using a 5-point Likert scale, with a 2-month wash-out period.

View Article and Find Full Text PDF

Photon-counting computed tomography (PCCT) is an innovative technological advancement in relation to x-ray detectors which offers ultra-high-resolution images. The current study aimed to evaluate the visualization ability of PCCT compared to cone-beam computed tomographic (CBCT) devices for challenging endodontic diagnostic tasks. A reference image of an anthropomorphic phantom was acquired using an industrial micro-CT device.

View Article and Find Full Text PDF

Objective: To present patient dose levels for different CBCT scanners, acquired by a dose monitoring tool in a University Hospital, as a function of field of view (FOV), operation mode, and patient age.

Methods: An integrated dose monitoring tool was used to collect radiation exposure data [type of CBCT unit, dose-area product (DAP), FOV size, and operation mode] and patient demographic information (age, referral department) performed on a 3D Accuitomo 170 and a Newtom VGI EVO unit. Effective dose conversion factors were calculated and implemented into the dose monitoring system.

View Article and Find Full Text PDF

Purpose: To assess if single shot acquisitions with solid-state dosimeters as well as Robson's method could replace ionization chambers for tube output and HVL measurements, saving medical physicists time.

Material And Methods: The energy responses of 4 solid-state dosimeters with automatic calculation of HVL were compared to ionization chamber measurements. Five anode/filter combinations were tested: Mo/Mo, Mo/Rh, Rh/Rh, W/Rh and W/Ag, from 24kVp to 35kVp.

View Article and Find Full Text PDF

The relevance of presampling modulation transfer function (MTF) measurements in digital mammography (DM) quality control (QC) is examined. Two studies are presented: a case study on the impact of a reduction in MTF on the technical image quality score and analysis of the robustness of routine QC MTF measurements. In the first study, two needle computed radiography (CR) plates with identical sensitivities were used with differences in the 50% point of the MTF ( ) larger than the limiting value in the European guidelines ( change between successive measurements).

View Article and Find Full Text PDF

Purpose: To evaluate image quality of chest radiography for a number of systems in Belgium, using a contrast-detail (c-d) test object and Visual Grading Analysis (VGA) of an anthropomorphic phantom.

Methods: The study comprised 22 chest imaging systems in Belgium. C-d data were measured using Leeds TO20 test object, imaged using poly(methyl methacrylate) (PMMA) thicknesses of 9, 13 and 16 cm.

View Article and Find Full Text PDF