Gas-phase ion/molecule reactions have been used extensively for the structural elucidation of organic compounds in tandem mass spectrometry. Reagents for ion/molecule reactions can be introduced into a mass spectrometer via a continuous flow apparatus or through a pulsed inlet system. However, most of these approaches enable the use of only a single reagent at a time.
View Article and Find Full Text PDFIsomeric O- and N-glucuronides are common drug metabolites produced in phase II of drug metabolism. Distinguishing these isomers by using common analytical techniques has proven challenging. A tandem mass spectrometric method based on gas-phase ion/molecule reactions of deprotonated glucuronide drug metabolites with trichlorosilane (HSiCl) in a linear quadrupole ion trap mass spectrometer is reported here to readily enable differentiation of the O- and N-isomers.
View Article and Find Full Text PDFTandem mass spectrometry based on ion-molecule reactions has emerged as a powerful tool for structural elucidation of ionized analytes. However, most currently used reagents were designed to react with protonated analytes, making them suboptimal for acidic analytes that are preferentially detected in negative ion mode. In this work we demonstrate that the phenoxide, carboxylate, and phosphate functionalities can be identified in deprotonated molecules by use of a combination of two reagents, diethylmethoxyborane (DEMB) and water.
View Article and Find Full Text PDFRationale: The oxidation of sulfur atoms is an important biotransformation pathway for many sulfur-containing drugs. In order to rapidly identify the sulfone functionality in drug metabolites, a tandem mass spectrometric method based on ion-molecule reactions was developed.
Methods: A phosphorus-containing reagent, trimethyl phosphite (TMP), was allowed to react with protonated analytes with various functionalities in a linear quadrupole ion trap mass spectrometer.
Rationale: N-Monosubstituted hydroxylamines correspond to an important class of metabolites for many bioactive molecules. In this study, a tandem mass spectrometric method based on ion/molecule reactions was developed for the identification of compounds with the N-monosubstituted hydroxylamino functionality.
Methods: The diagnostic ion/molecule reaction occurs between protonated analytes with 2-methoxypropene (MOP) inside a linear quadrupole ion trap mass spectrometer.