Purpose: The aim of this study was to test a novel dual-modality imaging system that combines full-field digital mammography (FFDM) and automated breast ultrasound (ABUS) in a single platform. Our Aceso system, named after the Greek goddess of healing, was specifically designed for the early detection of cancer in women with dense breast tissue.
Materials And Methods: Aceso was first tested using two industry standards: a Contrast Detail Mammography (CDMAM) phantom as endorsed by European Reference Organisation for Quality Assured Breast Screening and Diagnostic Services was used to assess the FFDM images; and the CIRS 040GSE ultrasound phantom was imaged to evaluate the quality of the ABUS images.