The PediVAS blood pump is a magnetically levitated centrifugal pump designed for pediatric bridge-to-decision or bridge-to-recovery in pediatric patients from 3-20kg in weight. In preparation for submission of an investigational device exemption (IDE) application, we completed a final six-animal series of pre-clinical studies. The studies were conducted under controlled conditions as prescribed by the recently released FDA guidance document for animal studies for cardiovascular devices.
View Article and Find Full Text PDFThe Levitronix PediVAS is an extracorporeal magnetically levitated pediatric ventricular assist system with an optimal flow rate range of 0.3-1.5 L/min.
View Article and Find Full Text PDFThe treatment of children with life-threatening cardiac and cardiopulmonary failure is a large and underappreciated public health concern. We have previously shown that the CentriMag is a magnetically levitated centrifugal pump system, having the utility for treating adults and large children (1,500 utilized worldwide). We present here the PediVAS, a pump system whose design was modified from the CentriMag to meet the physiological requirements of young pediatric and neonatal patients.
View Article and Find Full Text PDF