Aim: Before clinical implementation of an approved electromagnetic tracking system (CAPPA IRAD EMT) an experimental trial was performed to investigate the accuracy of the system and its safety in application for transpedicular vertebral punctures in comparison to the classical fluoroscopic method.
Material And Methods: A total of 110 transpedicular punctures were performed bilaterally using 11 vertebrae of 5 realistic artificial phantoms and 1 pedicle was punctured with the conventional technique using c-arm fluoroscopy and the other with the electromagnetic tracking system. As a target a radiopaque non-ferromagnetic marker was implanted bilaterally in the anterior wall of the vertebrae.
Purpose: Percutaneous spine procedures may occasionally be difficult and subject to complications. Navigation using a dynamic reference base (DRB) may ease the procedure. Yet, besides other shortcomings, its fixation demands additional incisions and thereby defies the percutaneous character of the procedure.
View Article and Find Full Text PDF