High-temperature calibration methods in additive manufacturing involve the use of advanced techniques to accurately measure and control the temperature of the build material during the additive manufacturing process. Infrared cameras, blackbody radiation sources and non-linear optimization algorithms are used to correlate the temperature of the material with its emitted thermal radiation. This is essential for ensuring the quality and repeatability of the final product.
View Article and Find Full Text PDFAn improved apparatus for measuring the spectral directional emissivity in the wavelength range between 1 µm and 20 µm at temperatures up to 2400 K is presented in this paper. As a heating unit an inductor is used to warm up the specimen, as well as the blackbody reference to the specified temperatures. The heating unit is placed in a double-walled vacuum vessel.