Background: The purpose of this study was to describe the use of the VIPER software for patient-specific quality assurance (PSQA) of single-isocenter multitarget (SIMT) stereotactic radiosurgery (SRS) plans.

Materials And Methods: Twenty clinical of intensity-modulated (IMRT) SIMT SRS plans were reviewed. A total of 88 brain metastases were included.

View Article and Find Full Text PDF

Purpose: This study aimed to describe the feasibility of the online adaptive radiation therapy (oART) method developed at the Hospital Quirónsalud Barcelona for prostate cancer, using a standard C-arm linear accelerator (linac) and without the support of artificial intelligence.

Methods And Materials: The first 18 patients treated at the Hospital Quirónsalud Barcelona with the developed oART method were included. An ultrahypofractionated radiation therapy scheme consisting of 7 × 6.

View Article and Find Full Text PDF
Article Synopsis
  • The study evaluates the accuracy of the VIPER software, which reconstructs a 3D dose distribution for radiosurgery, by comparing it with a standard Monte Carlo simulation.
  • It involves testing several treatment plans using a 6 MV photon beam and assessing the results with ionization chambers and radiochromic films for accurate dose measurements.
  • Findings indicate that VIPER is a reliable tool for verifying multiple-target stereotactic radiosurgery plans, achieving high gamma passing rates and requiring no setup time on the linear accelerator (linac).
View Article and Find Full Text PDF

Aim: To investigate whether the target-isocenter deviations reported by a conventional Winston-Lutz (WL) test actually reflect the shifts of the measured prescription isodose line with respect to the target.

Background: A conventional WL test uses a metallic ball as a target that aims at several fields. But this test does not report information on the accuracy of the delivery in terms of dose.

View Article and Find Full Text PDF

Purpose: To evaluate the spatial accuracy of a frameless cone-beam computed tomography (CBCT)-guided cranial radiosurgery (SRS) using an end-to-end (E2E) phantom test methodology.

Methods And Materials: Five clinical SRS plans were mapped to an acrylic phantom containing a radiochromic film. The resulting phantom-based plans (E2E plans) were delivered four times.

View Article and Find Full Text PDF

In the present study, we describe a method based on the analysis of the dynamic MLC log files (DynaLog) generated by the controller of a Varian linear accelerator in order to perform patient-specific IMRT QA. The DynaLog files of a Varian Millennium MLC, recorded during an IMRT treatment, can be processed using a MATLAB-based code in order to generate the actual fluence for each beam and so recalculate the actual patient dose distribution using the Eclipse treatment planning system. The accuracy of the DynaLog-based dose reconstruction procedure was assessed by introducing ten intended errors to perturb the fluence of the beams of a reference plan such that ten subsequent erroneous plans were generated.

View Article and Find Full Text PDF