A Hadamard transform-capillary electrophoresis-UV (HT-CE-UV) detection technique is described for the analysis of biological samples. Pseudorandom injections of sample and buffer according to a simplex matrix obtained from the corresponding Hadamard matrix is performed with conventional capillaries. Alternating injections are achieved with a novel capillary "T" connector created by drilling conventional capillary dimensions through a 1-cm diameter polycarbonate disc.
View Article and Find Full Text PDFA Hadamard transform (HT) detection method for microchip capillary electrophoresis with laser-induced fluorescence and a charge-coupled device (CCD) is described and compared to signal-averaged detection. A low-noise CCD camera is used to image a section of a separation channel where each camera pixel can be thought of as a unique detector. For signal averaging, electropherograms corresponding to individual pixels can be averaged for improved S/N.
View Article and Find Full Text PDFThis paper describes an improved format for Shah convolution Fourier transform (SCOFT) detection that utilizes the spatial resolution of a charge-coupled device (CCD) rather than a fixed optical mask to perform a Shah or sine convolution over a fluorescence signal. The laser-induced fluorescence from a 9-mm section of microfabricated channel is collected with a CCD at 28 Hz. Each image frame is multiplied by a convolution function to modulate the collected signal through space.
View Article and Find Full Text PDF