Long-term field comparisons of continuous and integrated filter measurements of mass concentrations of particulate matter (PM) with an aerodynamic diameter less than or equal to 2.5 microm (PM2.5) were performed at rural and urban sites in New York State.
View Article and Find Full Text PDFJ Air Waste Manag Assoc
October 2004
Laboratory tests with generated aerosols were conducted to test the efficacy of two recent design modifications to the well-established tapered element oscillating microbalance (TEOM) continuous particulate matter (PM) mass monitor. The two systems tested were the sample equilibration system-equipped TEOM monitor operating at 30 degrees C, which uses a Nafion dryer as part of the sample inlet, and the differential TEOM monitor, which adds a switched electrostatic precipitator and uses a self-referencing algorithm to determine "true PM mass." Test aerosols included ammonium sulfate, ammonium nitrate, sodium chloride, copper (II) sulfate, and mixed aerosols.
View Article and Find Full Text PDF