The purpose of the present study was to test the behavior of two different generations of cochlear implant systems subjected to a clinical radiotherapy scheme and to determine the maximal acceptable cumulative radiation levels at which the devices show out-of-specification behaviors. Using stereotactic irradiation (Cyberknife, 6 MV photon beam), three Digisonic SP and three Neuro devices were submitted to 5 Gy doses that cumulated to 60 Gy (12 sessions) and 80 Gy (16 sessions), respectively. A follow-up series of irradiation was then applied, in which Digisonic SP devices received two additional fractions of 50 Gy each, cumulating to 160 Gy, and Neuro devices three additional fractions of 20, 40, and 150 Gy, cumulating to 290 Gy.