Purpose: To evaluate the accuracy and reproducibility of quantitative chemical shift-encoded (CSE) MRI to quantify proton-density fat-fraction (PDFF) in a fat-water phantom across sites, vendors, field strengths, and protocols.
Methods: Six sites (Philips, Siemens, and GE Healthcare) participated in this study. A phantom containing multiple vials with various oil/water suspensions (PDFF:0%-100%) was built, shipped to each site, and scanned at 1.
The loss of pulmonary artery (PA) compliance has significant pathophysiological effect on the right ventricle. Noninvasive and reliable assessment of PA wall stiffness would be an essential determiner of right heart load and a clinically useful factor to assess cardiovascular risk. Two MRI techniques have been proposed for assessing PA stiffness by measuring pulse wave velocity (PWV): transit time (TT) and flow area (QA).
View Article and Find Full Text PDFBackground: The assessment of arterial stiffness is increasingly used for evaluating patients with different cardiovascular diseases as the mechanical properties of major arteries are often altered. Aortic stiffness can be noninvasively estimated by measuring pulse wave velocity (PWV). Several methods have been proposed for measuring PWV using velocity-encoded cardiovascular magnetic resonance (CMR), including transit-time (TT), flow-area (QA), and cross-correlation (XC) methods.
View Article and Find Full Text PDF