Purpose: Cardiac positron emission tomography/magnetic resonance imaging (PET/MRI) acquisition presents novel clinical applications thanks to the combination of viability and metabolic imaging (PET) and functional and structural imaging (MRI). However, the resolution of PET, as well as cardiac and respiratory motion in nongated cardiac imaging acquisition protocols, leads to a reduction in image quality and severe quantitative bias. Respiratory or cardiac motion is customarily addressed with gated reconstruction which results in higher noise.
View Article and Find Full Text PDFUnlabelled: This study aimed to investigate image quality for a comprehensive set of isotopes ((18)F, (11)C, (89)Zr, (124)I, (68)Ga, and (90)Y) on 2 clinical scanners: a PET/CT scanner and a PET/MR scanner.
Methods: Image quality and spatial resolution were tested according to NU 2-2007 of the National Electrical Manufacturers Association. An image-quality phantom was used to measure contrast recovery, residual bias in a cold area, and background variability.