Publications by authors named "Jason Blonshine"

Background: The 2017 American Thoracic Society/European Respiratory Society (ATS/ERS) diffusing capacity of the lung for carbon monoxide (D) standards specify a control rule for assessing biologic quality control (BioQC) but have limited guidance on how to establish expected values for control rule variables. This study aimed to determine expected values for D BioQC using coefficient of variation (CV) and compare that the mean ± 2 SD control rule yields the same precision as mean ± 12% of the mean.

Methods: D BioQC data were collected from a multi-center inhaled medication study.

View Article and Find Full Text PDF

Background: Although quality control standards are recommended to ensure accurate test results, the coefficient of variation for the FVC and FEV biologic quality control (BioQC) is not specified. The primary aim of this study was to evaluate variations in spirometry BioQCs in a large and diverse cohort of individuals to determine an acceptable standard for the coefficient of variation.

Methods: The FVC and FEV biologic control data were secondary analyses from an inhaled medication trial that was conducted over 3 y ending in 2018 that included 114 laboratories.

View Article and Find Full Text PDF