We have proposed a long-term, noninvasive, nonrestrictive method of delivering and implanting a biosensor within the body via a swallowable implantation capsule robot (ICR). The design and preliminary validation of the ICR's primary subsystem-the sensor deployment system-is discussed and evidence is provided for major design choices. The purpose of the sensor deployment system is to adhere a small biosensor to the mucosa of the intestine long-term, and the modality was inspired by tapeworms and other organisms that employ a strategy of mechanical adhesion to soft tissue via the combined use of hooks or needles and suckers.