Publications by authors named "Jan Zajc"

Introduction: This review critically examines the efficacy of dual-modular stems in primary total hip arthroplasty. Given the variability and non-comparability of certain femoral stem designs and stem-neck couplings, with some even being withdrawn from the market, this review offers an in-depth analysis of predominant implant performances.

Areas Covered: The paper explores a brief historical summary related to dual-modular stems, including the complications associated with their use, diagnostic tools for evaluation, analysis of both recalled and currently available models, as well as alternative therapeutic options.

View Article and Find Full Text PDF

: Increased revision rate of dual-modular (DM) femoral stems in primary total hip arthroplasty (THA) because of modular-neck breakage and adverse local tissue reactions (ALTRs) to additional junction damage products is well established and some designs have been recalled from the market. However, some long-term studies of specific DM stems did not confirm the inferiority of these stems compared to standard single-modular (SM) stems, and a head-to-head comparison THA is missing. The objectives of this multicentre study were to determine the survivorship and complication rates of a common DM stem design compared to a similar SM stem.

View Article and Find Full Text PDF

Titanium (Ti) alloys have been proven to be one of the most suitable materials for orthopaedic implants. Dual modular stems have been introduced to primary total hip arthroplasty (THA) to enable better control of the femoral offset, leg length, and hip stability. This systematic review highlights information acquired for dual modular Ti stem complications published in the last 12 years and offers a conclusive discussion of the gathered knowledge.

View Article and Find Full Text PDF

Background: Bi-modular stems were introduced in primary total hip arthroplasty (THA) to enable better control of the femoral offset, leg length, and hip stability. Despite numerous reports on modular femoral neck fractures, some designs are still marketed worldwide. While the risk factors for the sudden failure are multifactorial and mostly known, the timing of this new THA complication is not predictable by any means.

View Article and Find Full Text PDF

Introduction: Revision of a well-fixed stem due to unexpected modular neck fracture is a catastrophe for the patient and a challenge for the surgeon. This study aimed to test the possibility of predicting interchangeable neck fracture from serum levels of the stem/neck alloy-consisting metals.

Materials And Methods: Nineteen patients at high risk for interchangeable neck fracture were randomly selected out of a cohort of 680 bimodular stems made from Ti6Al4V alloy.

View Article and Find Full Text PDF

The authors report on a case of modular femoral neck fracture which appeared 21 months after revision of acetabular component. The revision surgery was performed 8 years after the primary total hip arthroplasty due to aseptic loosening of the acetabular component. During acetabular revision, the primary implanted short (S, - 3.

View Article and Find Full Text PDF