Magnetisation Transfer Contrast (MTC) is an important MR contrast-generating mechanism to characterise the MR-invisible macromolecular protons using an off-resonance pre-saturation RF irradiation pulse (or MT pulse). MTC MRI is normally implemented at a fixed magnetic field; however, it may be useful to evaluate changes of the MT effect as a function of external magnetic field strength (B₀). In order to conduct field-dependent MTC experiments with a single MR system, two techniques are crucially needed.

View Article and Find Full Text PDF

We have investigated the design and construction of liquid nitrogen cooled surface coils made from stranded (litz) copper wire for low field MRI applications. If designed correctly, cooled litz coils can provide a competitive alternative to high temperature superconducting (HTS) coils without the complications associated with flux trapping. Litz coils can also be produced with a wider range of shapes and sizes, and at lower cost.

View Article and Find Full Text PDF

New improved pH-sensitive nitroxides were applied for in vivo studies. An increased stability of the probes towards reduction was achieved by the introduction of the bulky ethyl groups in the vicinity of the paramagnetic NO fragment. In addition, the range of pH sensitivity of the approach was extended by the synthesis of probes with two ionizable groups, and, therefore, with two pKa values.

View Article and Find Full Text PDF

The development of a 3-D, multi-nuclear continuous wave NMR imaging (CW-NMRI) system is described and its imaging capability is demonstrated on a range of materials exhibiting extremely short T(2) relaxation values. A variety of radiofrequency resonators were constructed and incorporated into a new gradient and field offset coil assembly, while the overall system design was modified to minimise microphonic noise which was present in an earlier prototype system. The chemically combined (27)Al in a high temperature refractory cement was imaged, and the CW-NMRI system was found to be sensitive to small differences in (27)Al content in these samples.

View Article and Find Full Text PDF

This paper describes the design, construction and use of a field-cycled proton-electron double-resonance imaging (FC-PEDRI) system for the detection and imaging of free radicals. The unique feature of this imager is its use of a 450-mT detection magnetic field in order to achieve good image quality and sensitivity. The detection magnetic field is provided by a superconducting magnet, giving high stability and homogeneity.

View Article and Find Full Text PDF

Two quantitative methods were developed for investigation of the potential of MRI for in vivo monitoring of retinal damage and detachment in experimental autoimmune uveitis (EAU). Measurements of retinal thickness and detachment area were performed on matched MR and histologic (HIST) images of rat eyes at different stages of EAU. In vivo MR images of rat eyes were acquired at 4.

View Article and Find Full Text PDF

A prototype continuous wave MRI system operating at 7T has been used successfully to study a variety of heterogeneous materials exhibiting T2 relaxation values ranging from 10 micros to 50 ms. Two-dimensional images of a poly(methly methacrylate) (PMMA) resolution phantom (T2=38 micros) exhibited a spatial resolution of approximately 1mm at a magnetic field gradient strength of 200 mT/m. The technique was used to study the hydration, drying, and subsequent water penetration properties of cement samples made from ordinary Portland cement, and revealed inhomogeneities arising from the cure conditions.

View Article and Find Full Text PDF

A study was made of the in vivo detectability of a pH-sensitive, imidazolidine spin probe, and the efficacy of low-frequency electron spin resonance (ESR)-based techniques for pH measurement in vitro and in vivo in rats. The techniques used were longitudinally-detected ESR (LODESR) and field-cycled dynamic nuclear polarization (FC-DNP) for in vitro and in vivo measurements, and radiofrequency (RF)- and X-band ESR for comparisons in vitro. The spin probe was hexamethyl imidazolidine (HMI) with a pK of 4.

View Article and Find Full Text PDF