The goal of this work is to present a systematic Monte Carlo validation study on the clinical implementation of the enhanced dynamic wedges (EDWs) into the Pinnacle(3) (Philips Medical Systems, Fitchburg, WI) treatment planning system (TPS) and QA procedures for patient plan verification treated with EDWs. Modeling of EDW beams in the Pinnacle(3) TPS, which employs a collapsed-cone convolution superposition (CCCS) dose model, was based on a combination of measured open-beam data and the 'Golden Segmented Treatment Table' (GSTT) provided by Varian for each photon beam energy. To validate EDW models, dose profiles of 6 and 10 MV photon beams from a Clinac 2100 C/D were measured in virtual water at depths from near-surface to 30 cm for a wide range of field sizes and wedge angles using the Profiler 2 (Sun Nuclear Corporation, Melbourne, FL) diode array system.
View Article and Find Full Text PDF