The verification of trace explosives detection systems is often constrained to small sample sets, so it is important to support the significance of the results with statistical analysis. As binary measurements, the trials are assessed using binomial statistics. A method is described based on the probability confidence interval and expressed in terms of the upper confidence interval bound that reports the probability of successful detection and its level of statistical confidence.
View Article and Find Full Text PDFThe vibrational bands of erythritol tetranitrate (ETN) were measured experimentally with both Raman spectroscopy and attenuated total reflectance Fourier transform infrared (ATR FT-IR) spectroscopy. Seventy-two (3N-6) vibrational modes were predicted for ETN using density functional theory calculations performed using the B3LYP/6-31G* density functional basis set and geometry optimization. Raman spectroscopy and ATR FT-IR were used to measure observable Raman and IR signatures between 140 and 3100 wavenumbers (cm(-1)).
View Article and Find Full Text PDFRaman spectral data collected with high-resolution laboratory spectrometers are processed into a format suitable for importing as a user library on a 1064 nm DeltaNu first generation, field-deployable spectrometer prototype. The two laboratory systems used are a 1064 nm Bruker Fourier transform (FT)-Raman spectrometer and a 785 nm Kaiser dispersive spectrometer. The steps taken to adapt for device-dependent spectral resolution, wavenumber shifts between instruments, and relative intensity response are described.
View Article and Find Full Text PDF