Publications by authors named "Jacob Hartsell"

Aim: This study evaluated the performance characteristics of a prototype Becton Dickinson (BD) (Franklin Lakes, NJ) glucose/galactose binding protein (GGBP) sensor placed intradermally (BD-ID) or subcutaneously (BD-SC) for continuous glucose monitoring.

Materials And Methods: The performance characteristics of the prototype BD GGBP sensor after intradermal or subcutaneous placement were assessed, and its accuracy was compared with that of a glucose oxidase (GOx)-based sensor and a standard laboratory method (YSI STAT2300 analyzer, Yellow Springs Instrument, Yellow Springs, OH) under glucose clamp conditions and during an off-clamp meal challenge in 40 patients with type 1 or 2 diabetes in a 12-h feasibility study.

Results: BD-ID and BD-SC sensors performed as well as or better than the GOx-based sensor (differences in median absolute percentage error 2-4 points in hyperglycemic and euglycemic regions, ≥ 10 points in the hypoglycemic region).