The filterscope diagnostic on DIII-D utilizes photomultiplier tubes to measure visible light emission from the plasma. The system has undergone a substantial upgrade since previous attempts to cross-calibrate the filterscope with other spectroscopic diagnostics were unsuccessful. The optics now utilize a dichroic mirror to initially split the light at nearly 99% transmission or reflectance for light below or above 550 nm.
View Article and Find Full Text PDFA multi-camera soft x-ray diagnostic has been developed to measure the equilibrium electron temperature profile and temperature fluctuations due to magnetohydrodynamic activity on the Compact Toroidal Hybrid experiment. The diagnostic consists of three separate cameras each employing two 20-channel diode arrays that view the same plasma region through different beryllium filter thicknesses of 1.8 μm and 3.