We report on the effect of lung-density correction on dose distribution in a transverse slice containing the isocenter for tangential-fields breast irradiation. In this case study we analyzed the target coverage as well as hot spots for four types of treatment plans: Plan 1 assumes uniform unit density throughout, Plan 2 utilizes all the treatment parameters of Plan 1, but takes into account the lower lung density. Plan 3 is generated by optimizing the dose distribution in the presence of the lower lung density, and Plan 4 is an improvement on Plan 3 by using custom instead of standard wedges.