Background: Implantable heart rhythm devices are susceptible to interference in hospitals where electromagnetic interference (EMI) sources are ubiquitous.
Case Description: We report three cases in which EMI from the external defibrillator caused the inability to interrogate Boston Scientific cardiac resynchronization therapy-pacemaker (CRT-P) devices. We have documented interference with the Boston Scientific CRT-P Contact Renewal device model numbers H120/H125 (Natick, MA, USA) and two brands of external defibrillators: the Philips Heartstart XL model number M4735A (Andover, MA, USA) and the Hewlett-Packard Codemaster model number M1722B (Palo Alto, CA, USA).