Six months after subcutaneous implantable cardioverter defibrillator (S-ICD) implantation a 26-year-old Brugada patient presented because of a beeping tone emitted by his device. Chest X-ray displayed two functionless transvenous shock leads and the S-ICD system with a lead fracture. During lead revision procedure, extensive preparation of the lead from unexpectedly firm surrounding fibrous tissue encapsulating the lead was necessary before it could be removed, and a new shock lead could be implanted.