Bench tests for 38 models of artificial heart valves (AHV) with different design parameters allowed us to decide in favour of the valves with reduced eccentricity (compared to the serial AHV of the EMAHV type) according to its resistance in the constant flow. Out of the compatibility checks of the design parameters tested it was concluded that the disk did not make the complete calculated angle when rotated. The dependence of AHV resistance on the disk rotation angle showed that there is no necessity to increase that angle more than 70 degrees for the mitral valve and more than 75 degrees for the aortic AHV.
View Article and Find Full Text PDFVestn Akad Med Nauk SSSR
January 1990
The present work is devoted to the clinical assessment of the Soviet-made cardiac valve prostheses 'EMIKS' and 'LIKS'. From 1983 to May 1, 1988, 632 prostheses were implanted to 508 patients. A group of 348 patients were assessed: 139 after mitral (M) replacement, 130 after aortal (A) replacement, and 79 after mitral-aortal (M+A) replacement.
View Article and Find Full Text PDFThe described method makes it possible to determine the kinematic parameters of the artificial cardial disk valves in treadmill tests. This method is more convenient and accurate than those proposed earlier. The article presents design formulae and graphs for determining the above parameters.
View Article and Find Full Text PDF