Purpose: This study investigated the use of high spatial resolution solid-state detectors (DUO and Octa) combined with an inclinometer for machine-based quality assurance (QA) of Volumetric Modulated Arc Therapy (VMAT) with flattened and flattening filter-free beams.

Method: The proposed system was inserted in the accessory tray of the gantry head of a Varian 21iX Clinac and a Truebeam linear accelerator. Mutual dependence of the dose rate (DR) and gantry speed (GS) was assessed using the standard Varian customer acceptance plan (CAP).

View Article and Find Full Text PDF

The multichannel vaginal cylinder (MVC) applicator employed for gynecological high dose rate (HDR) brachytherapy increases dose delivery complexity, and thus makes the treatment more prone to errors. A quality assurance (QA) procedure tracking the source throughout dose delivery can detect dwell position and time errors in the multiple channels of the applicator. A new MVC system with integrated real time in vivo treatment delivery QA has been developed based on diodes embedded on the outer surface of the MVC.

View Article and Find Full Text PDF

Purpose: In this work, the "edgeless" silicon detector technology is investigated, in combination with an innovative packaging solution, to manufacture silicon detectors with negligible angular response. The new diode is also characterized as a dosimeter for radiotherapy with the aim to verify its suitability as a single detector for in vivo dosimetry as well as large area 2D array that does not require angular correction to their response.

Methods: For the characterisation of the "edgeless-drop-in" detector technology, a set of samples have been manufactured with different sensitive areas (1 × 1 and 0.

View Article and Find Full Text PDF

Purpose: Spatial and temporal resolutions are two of the most important features for quality assurance instrumentation of motion adaptive radiotherapy modalities. The goal of this work is to characterize the performance of the 2D high spatial resolution monolithic silicon diode array named "MagicPlate-512" for quality assurance of stereotactic body radiation therapy (SBRT) and stereotactic radiosurgery (SRS) combined with a dynamic multileaf collimator (MLC) tracking technique for motion compensation.

Methods: MagicPlate-512 is used in combination with the movable platform HexaMotion and a research version of radiofrequency tracking system Calypso driving MLC tracking software.

View Article and Find Full Text PDF

Purpose: High dose rate (HDR) brachytherapy is a treatment method that is used increasingly worldwide. The development of a sound quality assurance program for the verification of treatment deliveries can be challenging due to the high source activity utilized and the need for precise measurements of dwell positions and times. This paper describes the application of a novel phantom, based on a 2D 11 × 11 diode array detection system, named “magic phantom” (MPh), to accurately measure plan dwell positions and times, compare them directly to the treatment plan, determine errors in treatment delivery, and calculate absorbed dose.