Tetrahedron beam computed tomography (TBCT) performs volumetric imaging using a stack of fan beams generated by a multiple pixel X-ray source. While the TBCT system was designed to overcome the scatter and detector issues faced by cone beam computed tomography (CBCT), it still suffers the same large cone angle artifacts as CBCT due to the use of approximate reconstruction algorithms. It has been shown that iterative reconstruction algorithms are better able to model irregular system geometries and that algebraic iterative algorithms in particular have been able to reduce cone artifacts appearing at large cone angles.
View Article and Find Full Text PDFPurpose: To improve the image contrast-to-noise (CNR) ratio for low-Z target megavoltage cone-beam CT (MV CBCT) using a statistical projection noise suppression algorithm based on the penalized weighted least-squares (PWLS) criterion.
Methods: Projection images of a contrast phantom, a CatPhan(®) 600 phantom and a head phantom were acquired by a Varian 2100EX LINAC with a low-Z (Al) target and low energy x-ray beam (2.5 MeV) at a low-dose level and at a high-dose level.
Purpose: A statistical projection restoration algorithm based on the penalized weighted least-squares (PWLS) criterion can substantially improve the image quality of low-dose CBCT images. The performance of PWLS is largely dependent on the choice of the penalty parameter. Previously, the penalty parameter was chosen empirically by trial and error.
View Article and Find Full Text PDFPurpose: The authors propose a combined scatter reduction and correction method to improve image quality in cone-beam computed tomography (CBCT). Although using a beam-block approach similar to previous techniques to measure the scatter, this method differs in that the authors utilize partially blocked projection data obtained during scatter measurement for CBCT image reconstruction. This study aims to evaluate the feasibility of the proposed approach.
View Article and Find Full Text PDFPurpose: To propose an attenuation integral digital imaging (AIDI) technique for the treatment portal verification of conventional and intensity-modulated radiotherapy (IMRT).
Methods: In AIDI technique, an open in air fluence image Io and a patient fluence image I were acquired under the same exposure. Then after doing the dark field correction for both the Io and I, the AIDI image was simply calculated as log(Io/I), which is the attenuation integral along the ray path from the x-ray source to a detector pixel element.
Purpose: Cone beam CT is increasingly being used for daily patient positioning verification during radiation therapy treatments. The daily use of CBCT could lead to accumulated patient doses higher than the older technique of weekly portal imaging. There have been several studies focusing on measurement or calculation of the patient dose from CBCT recently.
View Article and Find Full Text PDFA positioning QA procedure for Varian's 2D/2D (kV/MV) and 3D/3D (planCT/CBCT) matching was developed. The procedure was to check: (1) the coincidence of on-board imager (OBI), portal imager (PI), and cone beam CT (CBCT)'s isocenters (digital graticules) to a linac's isocenter (to a pre-specified accuracy); (2) that the positioning difference detected by 2D/2D (kV/MV) and 3D/3D(planCT/CBCT) matching can be reliably transferred to couch motion. A cube phantom with a 2 mm metal ball (bb) at the center was used.
View Article and Find Full Text PDFPhys Med Biol
October 2009
This study is to evaluate the dose calculation accuracy using Varian's cone-beam CT (CBCT) for pelvic adaptive radiotherapy. We first calibrated the Hounsfield Unit (HU) to electron density (ED) for CBCT using a mini CT QC phantom embedded into an IMRT QA phantom. We then used a Catphan 500 with an annulus around it to check the calibration.
View Article and Find Full Text PDFIntroduction: Although flat-panel based X-ray imaging has been well tested in diagnostic radiology, its use as an image-guided-radiotherapy (IGRT) system in a treatment room is new and requires systematic assessment.
Materials And Methods: BrainLab Novalis IGRT system was used for this study. It consists of two floor mounted kV X-ray tubes projecting obliquely into two flat-panel detectors mounted on the ceiling.
With the increased use of cone beam CT (CBCT) for daily patient setup, the accumulated dose from CBCT may be significantly higher than that from simulation CT or portal imaging. The objective of this work is to measure the dose from daily pelvic scans with fixed technical settings and collimations. CBCT scans were acquired in half-fan mode using a half bowtie and x-rays were delivered in pulsed-fluoro mode.
View Article and Find Full Text PDFTo develop a quality assurance (QA) program for the On-Board Imager (OBI) system and to summarize the results of these QA tests over extended periods from multiple institutions. Both the radiographic and cone-beam computed tomography (CBCT) mode of operation have been evaluated. The QA programs from four institutions have been combined to generate a series of tests for evaluating the performance of the On-Board Imager.
View Article and Find Full Text PDFThe technologies with kilovoltage (kV) and megavoltage (MV) imaging in the treatment room are now available for image-guided radiation therapy to improve patient setup and target localization accuracy. However, development of strategies to efficiently and effectively implement these technologies for patient treatment remains challenging. This study proposed an aggregated technique for on-board CT reconstruction using combination of kV and MV beam projections to improve the data acquisition efficiency and image quality.
View Article and Find Full Text PDFAn artificial intelligence (AI)-guided inverse planning system was developed to optimize the combination of parameters in the objective function for intensity-modulated radiation therapy (IMRT). In this system, the empirical knowledge of inverse planning was formulated with fuzzy if-then rules, which then guide the parameter modification based on the on-line calculated dose. Three kinds of parameters (weighting factor, dose specification, and dose prescription) were automatically modified using the fuzzy inference system (FIS).
View Article and Find Full Text PDFA fuzzy logic technique was applied to optimize the weighting factors in the objective function of an inverse treatment planning system for intensity-modulated radiation therapy (IMRT). Based on this technique, the optimization of weighting factors is guided by the fuzzy rules while the intensity spectrum is optimized by a fast-monotonic-descent method. The resultant fuzzy logic guided inverse planning system is capable of finding the optimal combination of weighting factors for different anatomical structures involved in treatment planning.
View Article and Find Full Text PDFWe report an investigation on the accuracy of inhomogeneity correction in photon radiotherapy from CT scans with different settings. Specifically, the dosimetric differences from different CT scan parameters (kV, mAs) to phantoms and from different Hounsfield unit versus electron density (HU-ED) curves to patients are investigated. The absolute dose per monitor units (dose/MU) is used to quantify the results.
View Article and Find Full Text PDF