Publications by authors named "Henner Huflage"

Aim was to evaluate the influence of different quantum iterative reconstruction (QIR) levels on the image quality of femoral photon-counting CT angiographies (PCD-CTA).Ultra-high resolution PCD-CTA were obtained from both extremities of five extracorporeally-perfused cadavers using constant tube voltage and maximum radiation dose (71.2 ± 11.

View Article and Find Full Text PDF
Article Synopsis
  • The TFCC (Triangular Fibrocartilage Complex) has several components that have been better understood recently, and current classifications for it are somewhat lacking.
  • The review focuses on the anatomy, types of lesions visible in MRI, and compares these with existing literature, introducing the new CUP classification and illustrating it with images.
  • Accurate imaging requires advanced techniques like gadolinium-enhanced MRI or MR/CT arthrography, which are necessary for identifying the minute structures of the TFCC, especially the important stabilizing lamina fovealis.
View Article and Find Full Text PDF

Rationale And Objectives: This study investigates the dose burden of photon-counting detector (PCD) lung CT with ultra-high-resolution (UHR) and standard mode using organ-based tube current modulation (OBTCM).

Materials And Methods: An anthropomorphic Alderson-Rando phantom was scanned in UHR and standard mode with and without OBTCM on three dose levels (IQ 5, 20, 50). Effective radiation dose was determined by thermoluminescent dosimetry in 13 measurement sites and compared with the calculated effective dose derived from the dose-length product.

View Article and Find Full Text PDF

Background And Aims: The aim of this study was to investigate the imaging performance and quality differences of PCD-CT in standard resolution mode (SR) versus ultra-high resolution mode (UHR) in the lower extremity runoff of dose-matched CTAs in a human cadaveric model.

Methods: Extracorporeal perfusion of the upper leg was established in one fresh-frozen human cadaver via inguinal and popliteal accesses using a peristaltic pump. Seven peripheral stents were deployed in the SFA.

View Article and Find Full Text PDF

Photon-counting detectors (PCDs) have emerged as one of the most influential technical developments for medical imaging in recent memory. Surpassing conventional systems with energy-integrating detector technology in many aspects, PCD-CT scanners provide superior spatial resolution and dose efficiency for all radiological subspecialities. Demanding detailed display of trabecular microarchitecture and extensive anatomical coverage frequently within the same scan, musculoskeletal (MSK) imaging in particular can be a beneficiary of PCD-CT's remarkable performance.

View Article and Find Full Text PDF

Background Contrast-unenhanced abdominal CT is the imaging standard for urinary calculi detection; however, studies comparing photon-counting detector (PCD) CT and energy-integrating detector (EID) CT dose-reduction potentials are lacking. Purpose To compare the radiation dose and image quality of optimized EID CT with those of an experimental PCD CT scan protocol including tin prefiltration in patients suspected of having urinary calculi. Materials and Methods This retrospective single-center study included patients who underwent unenhanced abdominal PCD CT or EID CT for suspected urinary caliculi between February 2022 and March 2023.

View Article and Find Full Text PDF

Rationale And Objectives: To assess image quality and radiation dose of ultra-high-pitch CT pulmonary angiography (CTPA) with free-breathing technique for diagnosis of pulmonary embolism using a photon-counting detector (PCD) CT compared to matched energy-integrating detector (EID)-based single-energy CTPA.

Materials And Methods: Fifty-one PCD-CTPAs were prospectively compared to 51 CTPAs on a third-generation dual-source EID-CT. CTPAs were acquired with an ultra-high-pitch protocol with free-breathing technique (40 mL contrast medium, pitch 3.

View Article and Find Full Text PDF

Since the emergence of the first photon-counting computed tomography (PCCT) system in late 2021, its advantages and a wide range of applications in all fields of radiology have been demonstrated. Compared to standard energy-integrating detector-CT, PCCT allows for superior geometric dose efficiency in every examination. While this aspect by itself is groundbreaking, the advantages do not stop there.

View Article and Find Full Text PDF

Introduction: Cochlear duct length (CDL) measurement plays a role in the context of individualized cochlear implant (CI) surgery regarding an individualized selection and implantation of the CI electrode carrier and an efficient postoperative anatomy-based fitting process. The level of detail of the preoperative temporal bone CT scan depends on the imaging modality with major impact on CDL measurements and CI electrode contact position determination. The aim of this study was to evaluate the accuracy of perioperative CDL measurements and electrode contact determination in photon-counting CT (PCCT).

View Article and Find Full Text PDF

Rationale And Objectives: This study investigates the performance of tomosynthesis in the presence of osteosynthetic implants, aiming to overcome superimposition-induced limitations in conventional radiograms.

Materials And Methods: After surgical fracture induction and subsequent osteosynthesis, 8 cadaveric fracture models (wrist, metacarpus, ankle, metatarsus) were scanned with the prototypical tomosynthesis mode of a multiuse x-ray system. Tomosynthesis protocols at 60, 80, and 116 kV (sweep angle 10°, 13 FPS) were compared with standard radiograms.

View Article and Find Full Text PDF
Article Synopsis
  • The study aimed to analyze how helical pitch and gantry rotation time affect image quality and file size in ultrahigh-resolution photon-counting CT scans of cadaveric specimens.
  • Scans were conducted with varying pitches and rotation times, revealing that while scan duration differed significantly, image quality decreased with high-pitch and short rotation protocols despite consistent dose measures.
  • Findings indicate that careful selection of pitch and rotation time is crucial for optimizing image quality in clinical practices, as longer acquisition times lead to significantly larger raw data sizes, which can limit practical use.
View Article and Find Full Text PDF
Article Synopsis
  • Some medical scans called MRA take too long, so a new method called REACT was tested for quick images of blood vessels in stroke patients.
  • The study involved 76 stroke patients, and both REACT and another method (CE-MRA) were used to check for blockages in a main artery in the neck.
  • REACT showed very good results, finding 88.5% of serious blockages accurately and matching well with CE-MRA in how doctors rated the scans and their quality.
View Article and Find Full Text PDF

This study investigated whether virtual monoenergetic images (VMIs) and iodine mapping based on dual-energy CT (DECT) provide advantages in the assessment of endometrial cancer. A dual-source DECT was performed for primary staging of histologically proven endometrioid adenocarcinoma in 21 women (66.8 ± 12.

View Article and Find Full Text PDF

Rationale And Objectives: Aiming to offset image quality limitations in radiographs due to superimposition, this study investigates the diagnostic potential of appendicular skeleton tomosynthesis.

Materials And Methods: Eight cadaveric extremities (four hands and feet) were examined employing the prototypical tomosynthesis mode of a twin robotic X-ray scanner. 12 protocols with varying sweep angles (10, 20 vs.

View Article and Find Full Text PDF

Photon-counting detector (PCD)-CT allows for reconstruction of virtual non-iodine (VNI) images from contrast-enhanced datasets. This study assesses the diagnostic performance of aortic valve calcification scoring (AVCS) derived from VNI datasets generated with a 1st generation clinical dual-source PCD-CT. AVCS was evaluated in 123 patients (statistical analysis only comprising patients with aortic valve calcifications [n = 56; 63.

View Article and Find Full Text PDF

Rationale And Objectives: This study aims to compare the diagnostic confidence of photon-counting detector CT angiography (PCD-CTA) depending on the used vascular reformatting kernels with digital subtraction angiography (DSA) as diagnostic reference standard in peripheral arterial occlusive disease (PAOD).

Material And Methods: In 39 patients, 45 lower extremity PCD-CTA with subsequent DSA were analyzed. Advanced PAOD (Fontaine stage 4) was ascertained in 77.

View Article and Find Full Text PDF
Article Synopsis
  • The study investigates ways to reduce radiation doses in abdominal CT scans for diagnosing kidney stones, using spectral shaping and tube voltage modulation.
  • Eight cadaveric specimens were scanned at different tube voltages and doses, and the image quality was evaluated both quantitatively and subjectively by radiologists.
  • Results showed that using a tin filter at 100 kVp provided better image quality at lower doses compared to higher voltages and conventional imaging methods, indicating effective radiation dose reduction without compromising diagnostic quality.
View Article and Find Full Text PDF

Purpose: The aim of this study was to evaluate the usability of a recently developed extracorporeally-perfused cadaver model for training the angiographic management of acute arterial diseases and periprocedural complications that may occur during endovascular therapy of the lower extremity arterial runoff.

Materials And Methods: Continuous extracorporeal perfusion was established in three fresh-frozen body donors via inguinal and infragenicular access. Using digital subtraction angiography for guidance, both arterial embolization (e.

View Article and Find Full Text PDF

Objectives: Image acquisition in ultra-high-resolution (UHR) scan mode does not impose a dose penalty in photon-counting CT (PCCT). This study aims to investigate the dose saving potential of using UHR instead of standard-resolution PCCT for lumbar spine imaging.

Materials And Methods: Eight cadaveric specimens were examined with 7 dose levels (5-35 mGy) each in UHR (120 × 0.

View Article and Find Full Text PDF

Introduction: With increasing spatial resolution, diffusion-weighted imaging (DWI) may be suitable for morphologic lesion characterization in breast MRI - an area that has traditionally been occupied by dynamic contrast-enhanced imaging (DCE). This investigation compared DWI with b values of 800 and 1600 s/mm to DCE for lesion morphology assessment in high-resolution breast MRI at 3 Tesla.

Material And Methods: Multiparametric breast MRI was performed in 91 patients with 93 histopathologically proven lesions (31 benign, 62 malignant).

View Article and Find Full Text PDF

This retrospective study aims to provide an intra-individual comparison of aortic CT angiographies (CTAs) using first-generation photon-counting-detector CT (PCD-CT) and third-generation energy-integrating-detector CT (EID-CT). High-pitch CTAs were performed with both scanners and equal contrast-agent protocols. EID-CT employed automatic tube voltage selection (90/100 kVp) with reference tube current of 434/350 mAs, whereas multi-energy PCD-CT scans were generated with fixed tube voltage (120 kVp), image quality level of 64, and reconstructed as 55 keV monoenergetic images.

View Article and Find Full Text PDF

Background: With the emergence of photon-counting CT, ultrahigh-resolution (UHR) imaging can be performed without dose penalty. This study aims to directly compare the image quality of UHR and standard resolution (SR) scan mode in femoral artery angiographies.

Methods: After establishing continuous extracorporeal perfusion in four fresh-frozen cadaveric specimens, photon-counting CT angiographies were performed with a radiation dose of 5 mGy and tube voltage of 120 kV in both SR and UHR mode.

View Article and Find Full Text PDF

Purpose: To investigate the metal artifact suppression potential of combining tin prefiltration and virtual monoenergetic imaging (VMI) for osseous microarchitecture depiction in ultra-high-resolution (UHR) photon-counting CT (PCCT) of the lower extremity.

Method: Derived from tin-filtered UHR scans at 140 kVp, polychromatic datasets (T3D) and VMI reconstructions at 70, 110, 150, and 190 keV were compared in 117 patients with lower extremity metal implants (53 female; 62.1 ± 18.

View Article and Find Full Text PDF

Dual-energy CT pulmonary angiography (CTPA) with energy-integrating detector (EID) technology is limited by the inability to use high-pitch technique. The purpose of this study was to compare the image quality of anatomic images and iodine maps between high-pitch photon-counting detector (PCD) CTPA and dual-energy EID CTPA. This prospective study included 117 patients (70 men and 47 women; median age, 65 years) who underwent CTPA to evaluate for pulmonary embolism between March 2022 and November 2022.

View Article and Find Full Text PDF

Purpose: Since organ-based tube current modulation (OBTCM) and tin prefiltration are limited on their own in lowering the dose of lung CT examinations, this experimental study was designed to investigate whether combinations with anterior patient shielding can increase the dose reduction potential.

Material And Methods: Three pairs of scan protocols without/with breast shield (P1/P2: standard 120kVp, P3/P4: OBTCM at 100 kVp, P5/P6: Sn 100 kVp) were employed for radiation exposure and image quality comparisons on an anthropomorphic Alderson-Rando phantom. Equivalent doses were measured in eleven sites via thermoluminescent dosimetry and the effective dose was obtained by summation of the weighted organ doses.

View Article and Find Full Text PDF