Purpose: The development of a calibrationless parallel imaging method for accelerated simultaneous multi-slice (SMS) MRI based on Regularized Nonlinear Inversion (NLINV), evaluated using Cartesian and radial fast low-angle shot (FLASH).
Theory And Methods: NLINV is a parallel imaging method that jointly estimates image content and coil sensitivities using a Newton-type method with regularization. Here, NLINV is extended to SMS-NLINV for reconstruction and separation of all simultaneously acquired slices.