Past evaluation of particle contamination on packaged implants has typically been conducted using US Pharmacopeia (USP) 788, a 1970s pharmaceutical guideline created to evaluate contaminant particles in injectable fluids and syringes. Our objective was to reestablish relevant acceptance criteria for residual orthopedic and other implant debris, including smaller particles (i.e.
View Article and Find Full Text PDFBackground: This study aimed to assess metal sensitization ranges among orthopaedic patients by comparing adaptive immune responses in all-comer pre- and post-operative orthopaedic adults who were COVID-19 unvaccinated or vaccinated vs patients with a painful aseptic implant by lymphocyte transformation test (LTT) to SARS-CoV-2-Spike-Protein (SP) and implant metal(s), respectively.
Methods: Data were retrospectively reviewed from three independent groups: unvaccinated COVID-19 adults (n = 23); fully COVID-19 vaccinated adults (n = 35); unvaccinated, painful aseptic implant patients with history of metal allergy (n = 98). Standard in vitro LTT for SP and implant metal(s) (nickel, cobalt) were performed and rated as negative (stimulation index [SI]<2), mild (SI ≥ 2), positive (SI ≥ 4-15), and high sensitization (SI > 15) adaptive immune responses to tested antigen.
Metal wear and corrosion debris remain a limiting factor for long-term durability of total hip replacement (THR). Common wear particle production techniques for research differ from the actual tribocorrosion processes at the implant site, potentially causing loss of valuable information. The aim of this study was to investigate reactions to freshly generated and time-stabilized particles and ions released from CoCrMo-alloy using a bio-tribometer, which mimics conditions of the periprosthetic environment.
View Article and Find Full Text PDFIt is widely recognized that innate macrophage immune reactions to implant debris are central to the inflammatory responses that drive biologic implant failure over the long term. Less common, adaptive lymphocyte immune reactions to implant debris, such as delayed type hypersensitivity (DTH), can also affect implant performance. It is unknown which key patient factors, if any, mediate these adaptive immune responses that potentiate particle/macrophage mediated osteolysis.
View Article and Find Full Text PDFBackground: Although breast implants (BIs) have never been safer, factors such as implant debris may influence complications such as chronic inflammation and illness such as ALCL (anaplastic large cell lymphoma). Do different types of BIs produce differential particulate debris?
Objectives: The aim of this study was to quantify, investigate, and characterize the size, amount, and material type of both loosely bound and adherent surface particles on 5 different surface types of commercial BIs.
Methods: Surface particles from BIs of 5 surface types (n = 5/group), Biocell, Microcell, Siltex, Smooth, SmoothSilk, and Traditional-Smooth, were: (1) removed by a rinsing procedure and (2) removed with ultrapure adhesive carbon tabs.
Background: A battlefield-related injury results in increased local and systemic innate immune inflammatory responses, resulting in wound-specific complications and an increased incidence of osteoarthritis. However, little is known about whether severe injuries affect long-term systemic homeostasis, for example, immune function. Moreover, it also remains unknown whether battlefield-acquired metal fragments retained over the long term result in residual systemic effects such as altered immune reactivity to metals.
View Article and Find Full Text PDFThere has been increased interest in the role of corrosion in early implant failures and adverse local tissue reaction in total hip arthroplasty. We review the relationship between the different types of corrosion in orthopaedic surgery including uniform, pitting, crevice, and fretting or mechanically assisted crevice corrosion (MACC). Passive layer dynamics serves a critical role in each of these processes.
View Article and Find Full Text PDFWear and corrosion in total hip replacement negatively impact implant service-life and patient well-being. The aim of this study was to generate a statistical response surface of material loss using an apparatus, capable of testing the effect of wear and corrosion products in situ on cells, such as macrophages. The test chamber of a ball-on-flat tribometer operating inside a CO incubator was integrated with an electrochemical setup and adapted for cell culture work.
View Article and Find Full Text PDFCurrently, there is a dearth of information regarding the degree of particle shedding from breast implants (BIs) and what are the general biological consequences of BI debris. Thus, it is unclear to what degree BI debris compromises the long-term biological performance of BIs. For orthopedic implants, it is well established that the severity of biological reactivity to implant debris governs long-term clinical performance.
View Article and Find Full Text PDFInitiation of the innate sterile inflammatory response that can develop in response to microparticle exposure is little understood. Here, we report that a potent type 2 immune response associated with the accumulation of neutrophils, eosinophils and alternatively activated (M2) macrophages was observed in response to sterile microparticles similar in size to wear debris associated with prosthetic implants. Although elevations in interleukin-33 (IL-33) and type 2 cytokines occurred independently of caspase-1 inflammasome signalling, the response was dependent on Bruton's tyrosine kinase (BTK).
View Article and Find Full Text PDFMetal hypersensitivity has been recognized as an adverse biologic reaction that can compromise total joint arthroplasty (TJA) performance. However, the etiology of metal hypersensitivity responses in TJAs remains unclear. Metal implant debris is known to act as a danger signal that drives NLRP3 inflammasome activation.
View Article and Find Full Text PDFBackground: The rate of revision for some designs of total hip replacements due to idiopathic aseptic loosening has been reported as higher for women. However, whether this is environmental or inherently sex-related is not clear.
Objective: Can particle induced osteolysis be sex dependent? And if so, is this dependent on the type of implant debris (.
Options to treat early-onset scoliosis include guided-growth systems with sliding action between rods and pedicle screws. The wear was previously measured in an in vitro test, and in this in vivo rabbit model, we evaluated the local and systemic biological response to the stainless steel debris. Compared to the previous study, a relatively higher volume of representative wear particles with a median particle size of 0.
View Article and Find Full Text PDFJ Biomed Mater Res B Appl Biomater
November 2018
Posterior spinal fusion implants include number of interconnecting components, which are subjected to micromotion under physiological loading conditions inducing a potential for fretting corrosion. There is very little known about the fretting corrosion in these devices in terms of the minimum angular displacement (threshold) necessary to induce fretting corrosion or the amount of fretting corrosion that can arise during the life of the implant. Therefore, the first goal was to evaluate the threshold fretting corrosion in three anatomical orientations and second the long-term fretting corrosion for the three different material types of spinal implants under physiological loading conditions.
View Article and Find Full Text PDFBackground: The fretting-corrosion behavior of mixed metal contacts is affected by various mechanical and electrochemical parameters. Crevice conditions at the junction and patient-specific pathologies can affect the pH of the prosthetic environment. The main objective of this study is to understand the effect of pH variation at the stem/head junction of the hip implant under fretting corrosion exposure.
View Article and Find Full Text PDFBackground: Recent studies indicate that females demonstrate an increased risk of experiencing adverse local tissue reactions, aseptic loosening, and revision after primary metal-on-metal hip resurfacing arthroplasty compared with males; the underlying biological mechanisms responsible for sex discrepancies in implant failure remain unclear. In addition to anatomical and biomechanical sex differences, there may be inherent immunological disparities that predispose females to more aggressive adaptive immune reactivity to implant debris, i.e.
View Article and Find Full Text PDFFront Endocrinol (Lausanne)
January 2017
Despite the success in returning people to health saving mobility and high quality of life, the over 1 million total joint replacements implanted in the US each year are expected to eventually fail after approximately 15-25 years of use, due to slow progressive subtle inflammation to implant debris compromising the bone implant interface. This local inflammatory pseudo disease state is primarily caused by implant debris interaction with innate immune cells, i.e.
View Article and Find Full Text PDFCobalt alloy debris has been implicated as causative in the early failure of some designs of current total joint implants. The ability of implant debris to cause excessive inflammation via danger signaling (NLRP3 inflammasome) vs. pathogen associated pattern recognition receptors (e.
View Article and Find Full Text PDFInnate immune reactions to orthopedic implant debris are the primary cause of total joint replacement (TJR) failure over the long term (15-20 years). The role of pathogen associated pattern recognition receptors (i.e.
View Article and Find Full Text PDFJ Mech Behav Biomed Mater
September 2016
Background: Recently, there has been increasing concern in the orthopedic community over the use of hip implant modular devices due to an increasing number of reports of early failure, failure that has been attributed to fretting-corrosion at modular interfaces. Much is still unknown about the electrochemical and mechanical degradation mechanisms associated with the use of such devices.
Purpose: Accordingly, the purpose of our study was to develop a methodology for testing the fretting-corrosion behavior of modular junctions.
Spine (Phila Pa 1976)
April 2016
Almost 20% of joint replacement implants fail at 15 to 20 years. Reports suggest that systemic effects of metal-on-metal implants and local effects of total joint arthroplasty implants contributing to implant failure are immune system based. Sometimes implant wear debris can cause implant failure resulting from bone fracture, infection, or implant fracture/failure; most often, aseptic osteolysis or loosening leads to wear debris.
View Article and Find Full Text PDFPeriprosthetic osteolysis remains a major limitation of long-term successful total hip replacements with ultra-high molecular weight polyethylene (UHMWPE) bearings. As intra and extracellular reactive oxygen species are know to contribute to wear debris-induced osteoclastic bone resorption and decreased osteoblastic bone formation, antioxidant doped UHMWPE has emerged as an approach to reduce the osteolytic potential of wear debris and maintain coupled bone remodeling. To test this hypothesis in vivo, we evaluated the effects of crosslinked UHMWPE wear debris particles (AltrX(™) ), versus similar wear particles made from COVERNOX(™) containing UHMWPE (AOX(™) ), in an established murine calvaria model.
View Article and Find Full Text PDF